The quality of evidence for the analytical validation of sensor-based digital health technologies (sDHTs), which is the evaluation of algorithms converting sensor data into a clinically interpretable measure, is often inconsistent and insufficient. The existing V3+ framework codifies the overall evaluation process, which includes verification, usability validation, analytical validation, and clinical validation. To improve the scientific rigor of analytical validation, a hierarchical framework for selecting reference measures is needed because not all potential reference measures are of equal quality. The framework classifies reference measures based on attributes that contribute to reduced measurement variability, with defining and principal measures being the most rigorous due to objective data acquisition and the ability to retain source data.

Recommendations
The proposed framework sequentially moves the investigator through four steps: (1) Compile preliminary information, including the digital clinical measure, context of use (COU), algorithm requirements, and sensor verification evidence . (2) Select an existing reference measure, develop a novel comparator, or identify a set of anchor measures, prioritizing measures with the highest scientific rigor (defining → principal → manual → reported) . (3) Consider the impact of the data collection environment to determine if the analytical validation study can be conducted in the intended use environment with the highest-order measure, or if in-lab validation is necessary, ensuring the results are generalizable . (4) Describe the rationale for key study design decisions to encourage transparency for evaluators, regulators, and payers . Investigators must justify passing over a higher-ranked reference measure, generally only acceptable if the higher-ranked measure poses unacceptable risk or is not applicable to the context of use.

Regulatory Considerations
The principles of the framework for analytical validation apply regardless of the regulatory status of the sDHT (regulated medical device, low-risk general wellness apps, or research product) or its intended use (clinical care or clinical research). The framework is intended to help investigators support the most rigorous claims regarding sDHT performance, which is important for acceptance by evaluators, peer-reviewers, regulators, and payers. The categorization of the digital clinical measure as a digital biomarker or an electronic clinical outcome assessment also does not change the framework’s applicability.